This article has multiple issues. Please help improve it or discuss these issues on the talk page. (Learn how and when to remove these messages)
|
MINFLUX, or minimal fluorescence photon fluxes microscopy, is a super-resolution light microscopy method that images and tracks objects in two and three dimensions with single-digit nanometer resolution.[1][2][3]
MINFLUX uses a structured excitation beam with at least one intensity minimum – typically a doughnut-shaped beam with a central intensity zero – to elicit photon emission from a fluorophore. The position of the excitation beam is controlled with sub-nanometer precision, and when the intensity zero is positioned exactly on the fluorophore, the system records no emission. Thus, the system requires few emitted photons to determine the fluorophore's location with high precision. In practice, overlapping the intensity zero and the fluorophore would require a priori location knowledge to position the beam. As this is not the case, the excitation beam is moved around in a defined pattern to probe the emission from the fluorophore near the intensity minimum.[1]
Each localization takes less than 5 microseconds,[1] so MINFLUX can construct images of nanometric structures or track single molecules in fixed and live specimens by pooling the locations of fluorescent labels. Because the goal is to locate the point where a fluorophore stops emitting, MINFLUX significantly reduces the number of fluorescence photons needed for localization compared to other methods.[2][4]
A commercial MINFLUX system is available from abberior instruments GmbH.[5]
Principle
editMINFLUX overcomes the Abbe diffraction limit in light microscopy and distinguishes individual fluorescing molecules by leveraging the photophysical properties of fluorophores. The system temporarily silences (sets in an OFF-state) all but one molecule within a diffraction-limited area (DLA) and then locates that single active (in an ON-state) molecule.[1] Super-resolution microscopy techniques like stochastic optical reconstruction microscopy (STORM) and photoactivated localization microscopy (PALM) do the same.[6] However, MINFLUX differs in how it determines the molecule’s location.
The excitation beam used in MINFLUX has a local intensity minimum or intensity zero. The position of this intensity zero in a sample is adjusted via control electronics and actuators with sub-nanometer spatial and sub-microsecond temporal precision. When the active molecule located at is in a non-zero intensity area of the excitation beam, it fluoresces. The number of photons emitted by the active molecule is proportional to the excitation beam intensity at that position.
In the vicinity of the excitation beam intensity zero, the intensity of the emission from the active molecule when the intensity zero is located at position can be approximated by a quadratic function. Therefore, the recorded number of emission photons is:
where is a measure of the collection efficiency of detection, the absorption cross-section of the emitter, and the quantum yield of fluorescence.
In other words, photon fluxes emitted by the active molecule when it is located close to the zero-intensity point of the excitation beam carry information about its distance to the center of the beam. That information can be used to find the position of the active molecule. The position is probed with a set of excitation intensities . For example, the active molecule is excited with the same doughnut-shaped beam moved to different positions. The probing results in a corresponding set of photon counts . These photon counts are probabilistic; each time such a set is measured, the result is a different realization of photon numbers fluctuating around a mean value. Since their distribution follows Poissonian statistics, the expected position of the active molecule can be estimated from the photon numbers, using, for example, a maximum likelihood estimation of the form:
The position maximizes the likelihood that the measured set of photon counts occurred exactly as recorded and is thus, an estimate of the active molecule’s location.[7]
Localization process
editRecordings of the emitting active molecule at two different excitation beam positions are needed to use the quadratic approximation in the one-dimensional basic principle described above. Each recording provides a one-dimensional distance value to the center of the excitation beam. In two dimensions, at least three recording points are needed to ascertain a location that can be used to move the MINFLUX excitation beam toward the target molecule. These recording points demarcate a probing area L. Balzarotti et al.[1] use the Cramér-Rao limit to show that constricting this probing area significantly improves localization precision, more so than increasing the number of emitted photons:
where is the Cramér-Rao limit, is the diameter of the probing area, and is the number of emitted photons.
MINFLUX takes advantage of this feature when localizing an active fluorophore. It records photon fluxes using a probing scheme of at least three recording points around the probing area and one point at the center. These fluxes differ at each recording point as the active molecule is excited by different light intensities. Those flux patterns inform the repositioning of the probing area to center on the active molecule. Then the probing process is repeated. With each probing iteration, MINFLUX constricts the probing area , narrowing the space where the active molecule can be located. Thus, the distance remaining between the intensity zero and the active molecule is determined more precisely at each iteration. The steadily improving positional information minimizes the number of fluorescence photons and the time that MINFLUX needs to achieve precise localizations.[8]
Applications
editBy pooling the determined locations of multiple fluorescent molecules in a specimen, MINFLUX generates images of nanoscopic structures with a resolution of 1–3 nm.[9] MINFLUX has been used to image DNA origami[1][10] and the nuclear pore complex[11] and to elucidate the architecture of subcellular structures in mitochondria and photoreceptors.[12][13] Because MINFLUX does not collect large numbers of photons emitted from target molecules, localization is faster than with conventional camera-based systems.[14] Thus, MINFLUX can iteratively localize the same molecule at microsecond intervals over a defined period. MINFLUX has been used to track the movement of the motor protein kinesin-1, both in vitro and in vivo,[15][16] and to monitor configurational changes of the mechanosensitive ion channel PIEZO1.[17]
See also
edit- Confocal microscopy
- Fluorescence
- Fluorescence microscope
- Fluorescence resonance energy transfer microscopy
- Laser scanning confocal microscopy
- Optical microscopy
- Photoactivated localization microscopy
- Stochastic optical reconstruction microscopy
- Super-resolution microscopy
- Ground state depletion microscopy
- RESOLFT
References
edit- ^ a b c d e f Balzarotti, Francisco; Eilers, Yvan; Gwosch, Klaus C.; Gynnå, Arvid H.; Westphal, Volker; Stefani, Fernando D.; Elf, Johan; Hell, Stefan W. (2017-02-10). "Nanometer resolution imaging and tracking of fluorescent molecules with minimal photon fluxes". Science. 355 (6325): 606–612. arXiv:1611.03401. Bibcode:2017Sci...355..606B. doi:10.1126/science.aak9913. hdl:11858/00-001M-0000-002C-2D9A-4. ISSN 0036-8075. PMID 28008086. S2CID 5418707.
- ^ a b Sigal, Yaron M.; Zhou, Ruobo; Zhuang, Xiaowei (2018-08-31). "Visualizing and discovering cellular structures with super-resolution microscopy". Science. 361 (6405): 880–887. Bibcode:2018Sci...361..880S. doi:10.1126/science.aau1044. ISSN 0036-8075. PMC 6535400. PMID 30166485.
- ^ Tang, Verena (2023-01-10). "Ein Potenzial, das noch in keiner Form gehoben ist". Spektrum.de. Retrieved 2023-11-17.
- ^ Xiao, Jie; Ha, Taekjip (2017-02-10). "Flipping nanoscopy on its head". Science. 355 (6325): 582–584. Bibcode:2017Sci...355..582X. doi:10.1126/science.aam5409. ISSN 0036-8075. PMC 8989063. PMID 28183938.
- ^ "MINFLUX – @abberior.rocks".
- ^ Lelek, Mickaël; Gyparaki, Melina T.; Beliu, Gerti; Schueder, Florian; Griffié, Juliette; Manley, Suliana; Jungmann, Ralf; Sauer, Markus; Lakadamyali, Melike; Zimmer, Christophe (2021-06-03). "Single-molecule localization microscopy". Nature Reviews Methods Primers. 1 (1). doi:10.1038/s43586-021-00038-x. ISSN 2662-8449. PMC 9160414. PMID 35663461.
- ^ Eilers, Y. 2017. Localizing and tracking of fluorescent molecules with minimal photon fluxes. Doctoral Dissertation. Georg-August-University Göttingen. https://ediss.uni-goettingen.de/handle/11858/00-1735-0000-002E-E329-2 (accessed January 2024)
- ^ Gwosch, Klaus C.; Pape, Jasmin K.; Balzarotti, Francisco; Hoess, Philipp; Ellenberg, Jan; Ries, Jonas; Hell, Stefan W. (2020-02-02). "MINFLUX nanoscopy delivers 3D multicolor nanometer resolution in cells". Nature Methods. 17 (2): 217–224. doi:10.1038/s41592-019-0688-0. ISSN 1548-7091. PMID 31932776. S2CID 210168754.
- ^ Bond, Charles; Santiago-Ruiz, Adriana N.; Tang, Qing; Lakadamyali, Melike (2022-01-20). "Technological advances in super-resolution microscopy to study cellular processes". Molecular Cell. 82 (2): 315–332. doi:10.1016/j.molcel.2021.12.022. PMC 8852216. PMID 35063099.
- ^ Wogan, Tim (2017-01-09). "New super-resolution microscope combines Nobel-winning technologies". Physics World. Retrieved 2023-11-17.
- ^ Schmidt, Roman; Weihs, Tobias; Wurm, Christian A.; Jansen, Isabelle; Rehman, Jasmin; Sahl, Steffen J.; Hell, Stefan W. (2021-03-05). "MINFLUX nanometer-scale 3D imaging and microsecond-range tracking on a common fluorescence microscope". Nature Communications. 12 (1): 1478. Bibcode:2021NatCo..12.1478S. doi:10.1038/s41467-021-21652-z. ISSN 2041-1723. PMC 7935904. PMID 33674570.
- ^ Pape, Jasmin K.; Stephan, Till; Balzarotti, Francisco; Büchner, Rebecca; Lange, Felix; Riedel, Dietmar; Jakobs, Stefan; Hell, Stefan W. (2020-08-25). "Multicolor 3D MINFLUX nanoscopy of mitochondrial MICOS proteins". Proceedings of the National Academy of Sciences. 117 (34): 20607–20614. Bibcode:2020PNAS..11720607P. doi:10.1073/pnas.2009364117. ISSN 0027-8424. PMC 7456099. PMID 32788360.
- ^ Grabner, Chad P.; Jansen, Isabelle; Neef, Jakob; Weihs, Tobias; Schmidt, Roman; Riedel, Dietmar; Wurm, Christian A.; Moser, Tobias (2022-07-15). "Resolving the molecular architecture of the photoreceptor active zone with 3D-MINFLUX". Science Advances. 8 (28): eabl7560. Bibcode:2022SciA....8L7560G. doi:10.1126/sciadv.abl7560. ISSN 2375-2548. PMC 9286502. PMID 35857490.
- ^ Strack, Rita (2017-02-28). "Taking nanoscopy to the limit". Nature Methods. 14 (3): 221. doi:10.1038/nmeth.4211. ISSN 1548-7091.
- ^ Deguchi, Takahiro; Iwanski, Malina K.; Schentarra, Eva-Maria; Heidebrecht, Christopher; Schmidt, Lisa; Heck, Jennifer; Weihs, Tobias; Schnorrenberg, Sebastian; Hoess, Philipp; Liu, Sheng; Chevyreva, Veronika; Noh, Kyung-Min; Kapitein, Lukas C.; Ries, Jonas (2023-03-10). "Direct observation of motor protein stepping in living cells using MINFLUX". Science. 379 (6636): 1010–1015. Bibcode:2023Sci...379.1010D. doi:10.1126/science.ade2676. ISSN 0036-8075. PMC 7614483. PMID 36893247.
- ^ Dambeck, Susanne (2017-03-23). "New Super Tool for Cell Biology". Lindau Nobel Laureate Meetings. Retrieved 2024-01-22.
- ^ Mulhall, Eric M.; Gharpure, Anant; Lee, Rachel M.; Dubin, Adrienne E.; Aaron, Jesse S.; Marshall, Kara L.; Spencer, Kathryn R.; Reiche, Michael A.; Henderson, Scott C.; Chew, Teng-Leong; Patapoutian, Ardem (2023-08-31). "Direct observation of the conformational states of PIEZO1". Nature. 620 (7976): 1117–1125. Bibcode:2023Natur.620.1117M. doi:10.1038/s41586-023-06427-4. ISSN 0028-0836. PMC 10468401. PMID 37587339.