Sphygmomanometer

(Redirected from Sphygmomanometry)

A sphygmomanometer (/ˌsfɪɡmməˈnɒmɪtər/ SFIG-moh-mə-NO-mi-tər), also known as a blood pressure monitor, or blood pressure gauge, is a device used to measure blood pressure, composed of an inflatable cuff to collapse and then release the artery under the cuff in a controlled manner,[1] and a mercury or aneroid manometer to measure the pressure. Manual sphygmomanometers are used with a stethoscope when using the auscultatory technique.

A sphygmomanometer consists of an inflatable cuff, a measuring unit (the mercury manometer, or aneroid gauge), and a mechanism for inflation which may be a manually operated bulb and valve or a pump operated electrically.

Etymology

edit

The word sphygmomanometer uses the combining form of sphygmo- + manometer. The roots involved are as follows: Greek σφυγμός sphygmos "pulse", plus the scientific term manometer (from French manomètre), i.e. "pressure meter", itself coined from μανός manos "thin, sparse", and μέτρον metron "measure".[2][3][4]

Most sphygmomanometers were mechanical gauges with dial faces, or mercury columns, during most of the 20th century. Since the advent of electronic medical devices, names such as "meter" and "monitor" can also apply, as devices can automatically monitor blood pressure on an ongoing basis.

History

edit
 
A French sphygmomanometer used during World War I

The sphygmomanometer was invented by Samuel Siegfried Karl Ritter von Basch in the year 1881.[1] Scipione Riva-Rocci introduced a more easily-usable version in 1896. In 1901, pioneering neurosurgeon Dr. Harvey Cushing brought an example of Riva-Rocci's device to the US, modernized and popularized it within the medical community. Further improvement came in 1905 when Russian physician Nikolai Korotkov included diastolic blood pressure measurement following his discovery of "Korotkoff sounds". William A. Baum invented the Baumanometer brand in 1916,[5] while working for The Life Extension Institute which performed insurance and employment physicals.[6]

Types

edit

Both manual and digital meters are currently employed, with different trade-offs in accuracy versus convenience.

Manual

edit

A stethoscope is required for auscultation (see below). Manual meters are best used by trained practitioners, and, while it is possible to obtain a basic reading through palpation alone, this yields only the systolic pressure.

  • Mercury sphygmomanometers are considered the gold standard. They indicate pressure with a column of mercury, which does not require recalibration.[7] Because of their accuracy, they are often used in clinical trials of drugs and in clinical evaluations of high-risk patients, including pregnant women. A frequently used wall mounted mercury sphygmomanometer is also known as a Baumanometer.[6]
  • Aneroid sphygmomanometers (mechanical types with a dial) are in common use; they may require calibration checks, unlike mercury manometers. Aneroid sphygmomanometers are considered safer than mercury sphygmomanometers, although inexpensive ones are less accurate.[8] A major cause of departure from calibration is mechanical jarring. Aneroids mounted on walls or stands are not susceptible to this particular problem.

Digital

edit

Digital meters employ oscillometric measurements and electronic calculations rather than auscultation. They may use manual or automatic inflation, but both types are electronic, easy to operate without training, and can be used in noisy environments. They calculate systolic and diastolic pressures by oscillometric detection, employing either deformable membranes that are measured using differential capacitance, or differential piezoresistance, and they include a microprocessor.[9] They estimate mean arterial blood pressure and measure pulse rate; while systolic and diastolic pressures are obtained less accurately than with manual meters,[10] and calibration is also a concern.[11][12][13] Digital oscillometric monitors may not be advisable for some patients, such as those with arteriosclerosis, arrhythmia, preeclampsia, pulsus alternans, and pulsus paradoxus, as their calculations may not be correct for these conditions,[14][15] and in these cases, an analog sphygmomanometer is preferable when used by a trained person.

Digital instruments may use a cuff placed, in order of accuracy[16] and inverse order of portability and convenience, around the upper arm, the wrist, or a finger.[17] Recently, a group of researchers at Michigan State University developed a smartphone based device that uses oscillometry to estimate blood pressure.[18][19] The oscillometric method of detection used gives blood pressure readings that differ from those determined by auscultation, and vary according to many factors, such as pulse pressure, heart rate and arterial stiffness,[20] although some instruments are claimed also to measure arterial stiffness, and some can detect irregular heartbeats.

Operation

edit
 
Medical student taking blood pressure at the brachial artery
 
Explanation of how blood pressure is measured based on Korotkow sounds

In humans, the cuff is normally placed smoothly and snugly around an upper arm, at roughly the same vertical height as the heart while the subject is seated with the arm supported. Other sites of placement depend on species and may include the flipper or tail. It is essential that the correct size of cuff is selected for the patient. Too small a cuff results in too high a pressure, while too large a cuff results in too low a pressure. For clinical measurements it is usual to measure and record both arms in the initial consultation to determine if the pressure is significantly higher in one arm than the other. A difference of 10 mmHg may be a sign of coarctation of the aorta. If the arms read differently, the higher reading arm would be used for later readings.[21] The cuff is inflated until the artery is completely occluded.

With a manual instrument, listening with a stethoscope to the brachial artery, the examiner slowly releases the pressure in the cuff at a rate of approximately 2 mmHg per heart beat. As the pressure in the cuffs falls, a "whooshing" or pounding sound is heard (see Korotkoff sounds) when blood flow first starts again in the artery. The pressure at which this sound began is noted and recorded as the systolic blood pressure. The cuff pressure is further released until the sound can no longer be heard. This is recorded as the diastolic blood pressure. In noisy environments where auscultation is impossible (such as the scenes often encountered in emergency medicine), systolic blood pressure alone may be read by releasing the pressure until a radial pulse is palpated (felt). In veterinary medicine, auscultation is rarely of use, and palpation or visualization of pulse distal to the sphygmomanometer is used to detect systolic pressure.

Digital instruments use a cuff which may be placed, according to the instrument, around the upper arm, wrist, or a finger, in all cases elevated to the same height as the heart. They inflate the cuff and gradually reduce the pressure in the same way as a manual meter, and measure blood pressures by the oscillometric method.[9]

Significance

edit

By observing the mercury in the column, or the aneroid gauge pointer, while releasing the air pressure with a control valve, the operator notes the values of the blood pressure in mmHg. The peak pressure in the arteries during the cardiac cycle is the systolic pressure, and the lowest pressure (at the resting phase of the cardiac cycle) is the diastolic pressure. A stethoscope, applied lightly over the artery being measured, is used in the auscultatory method. Systolic pressure (first phase) is identified with the first of the continuous Korotkoff sounds. Diastolic pressure is identified at the moment the Korotkoff sounds disappear (fifth phase).

Measurement of the blood pressure is carried out in the diagnosis and treatment of hypertension (high blood pressure), and in many other healthcare scenarios.

See also

edit

References

edit
  1. ^ a b Booth J (1977). "A short history of blood pressure measurement". Proceedings of the Royal Society of Medicine. 70 (11): 793–9. doi:10.1177/003591577707001112. PMC 1543468. PMID 341169.
  2. ^ Harper D. "sphygmomanometer". Online Etymology Dictionary.
  3. ^ Harper D. "manometer". Online Etymology Dictionary.
  4. ^ σφυγμός, μανός, μέτρον. Liddell, Henry George; Scott, Robert; A Greek–English Lexicon at the Perseus Project.
  5. ^ U.S patent 1594039 Manometer
  6. ^ a b "Turning Mercury Into Solid Gold". The New York Times. 27 March 2005. Retrieved 5 July 2018.
  7. ^ "Comparing Mercury and Aneroid Sphygmomanometers". Sustainable Hospitals / Lowell Center for Sustainable Production. 2003. Retrieved 23 February 2015.
  8. ^ Misrin J. "Aneroid Sphygmomanometer: A Battle for Safer Blood Pressure Apparatus". Archived from the original on 10 August 2015. Retrieved 27 February 2012.
  9. ^ a b "Oscillometry, Explanation of oscillometric detection in Medical Electronics, N Townsend, p48-51" (PDF). Archived from the original (PDF) on 22 May 2013. Retrieved 10 February 2013.
  10. ^ "Oscillometric Method - Methods of Blood Pressure Measurement - Measurement of Blood PressureMethods of Blood Pressure Measurement -". www.severehypertension.net. 22 December 2007. Retrieved 13 April 2017.
  11. ^ Turner MJ. "Can we trust automatic sphygmomanometer validations?" Journal of Hypertension. 28(12), December 2010, pp. 2353–2356. doi:10.1097/HJH.0b013e32833e1011.
  12. ^ Turner MJ, van Schalkwyk JM (1 August 2008). "Automated Sphygmomanometers Should Not Replace Manual Ones, Based on Current Evidence". American Journal of Hypertension. 21 (8). Oxford University Press (OUP): 845. doi:10.1038/ajh.2008.204. ISSN 0895-7061. PMID 18648356.
  13. ^ Turner MJ1, Speechly C, Bignell N. (October 2007). "Sphygmomanometer calibration—why, how and how often?" Australian Family Physician. 36(10):834-838.
  14. ^ Hamzaoui O, Monnet X, Teboul JL (2013). "Pulsus paradoxus". Eur. Respir. J. 42 (6): 1696–705. doi:10.1183/09031936.00138912. PMID 23222878.
  15. ^ O'Brien E, Asmar R, Beilin L, Imai Y, Mallion JM, Mancia G, Mengden T, Myers M, Padfield P, Palatini P, Parati G, Pickering T, Redon J, Staessen J, Stergiou G, Verdecchia P (2003). "European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement". J. Hypertens. 21 (5): 821–48. doi:10.1097/00004872-200305000-00001. PMID 12714851. S2CID 3952069.
  16. ^ Mourad A, Gillies A, Carney S (2005). "Inaccuracy of wrist-cuff oscillometric blood pressure devices: an arm position artefact?" (PDF). Clinical Methods and Pathophysiology. 10 (2): 67–71. doi:10.1097/00126097-200504000-00003. PMID 15812253. S2CID 6100566.
  17. ^ "Blutdruckmessgerät - Handgelenk - Blutdruckmessgerät - Test" (in German). Blutdruckmessgeraet-vergleich-test.de. Retrieved 27 September 2016.
  18. ^ Chandrasekhar A (7 March 2018). "Smartphone-based blood pressure monitoring via the oscillometric finger-pressing method". Science Translational Medicine. 10 (431): eaap8674. doi:10.1126/scitranslmed.aap8674. PMC 6039119. PMID 29515001.
  19. ^ Chandrasekhar A (3 September 2018). "An iPhone Application for Blood Pressure Monitoring via the Oscillometric Finger Pressing Method". Scientific Reports. 8 (1): 13136. Bibcode:2018NatSR...813136C. doi:10.1038/s41598-018-31632-x. PMC 6120863. PMID 30177793.
  20. ^ van Montfrans GA (2001). "Oscillometric blood pressure measurement: progress and problems". Blood Press Monit. 6 (6): 287–90. doi:10.1097/00126097-200112000-00004. PMID 12055403.
  21. ^ Fred HL (2013). "Accurate Blood Pressure Measurements and the Other Arm". Texas Heart Institute Journal. 40 (3): 217–219. PMC 3709227. PMID 23914007.
edit
  • US patent 1089122, Francis Ashley Faught, Charles J Pilling, "Apparatus for measuring and indicating blood-pressure", issued 1914-03-03 
  • US patent 1594039, William A Baum, "Manometer", issued 1926-07-27 
  • US patent 2560237, R. H. Miller, "Sphygmomanometer", issued 1951-07-10 
  • US patent 6752764, Man S. Oh, "Pocket sphygmomanometer", issued 2004-06-22